New Generation of GC/High Resolution TOF-MS: APCI/APLI Source for Increased Flexibility
| Type: |
Application |
Application Note |
|
| Number: |
Technology |
LCGC0708 |
OTOF |
| Year |
Products |
2008 |
micrOTOF |
| |
Author |
| |
Thomas Arthen-Engeland and Ralf Dunsbach, Bruker Daltonik GmbH, Bremen, Germany |
| |
Reference |
| |
LCGC Europe THEAPPLICATIONSBOOK July 2008 |
| |
Abstract |
| |
TOF mass spectrometers have been widely used in combination with liquid chromatography. These systems can demonstrate a mass resolution of 15000 FWHM and a mass accuracy of 2 ppm. TOF systems that are designed to work with gas chromatography usually demonstrate a reduced performance. To date, a solution that uses the same TOF-MS on both GC and LC systems, merely by exchanging the chromatographic system, has not existed. Now, the newly developed Bruker Daltonics APCI source allows a micrOTOF, which is basically an LC–TOF-MS, to be coupled with a gas chromatograph. The performance of the micrOTOF, in terms of mass resolution and mass accuracy, is not influenced by the chromatographic system used for separation. Atmospheric pressure laser ionization (APLI) has been described recently as a useful alternative ionization method for certain applications where high selectivity and sensitivity is required.¹ APLI can also be realized with the new Bruker Daltonics source. |
| |